Mc-techgroup CSS-VC2 Operator's manual Uživatelský manuál Strana 24

  • Stažení
  • Přidat do mých příruček
  • Tisk
  • Strana
    / 56
  • Tabulka s obsahem
  • KNIHY
  • Hodnocené. / 5. Na základě hodnocení zákazníků
Zobrazit stránku 23
24 Gas sampling and gas conditioning technology 14.1a-ME
12.4.3 CONNECTION OF HEATED LINE FOR PORTABLE VERSION (ART. NO. 01G6190)
If, for the portable version in a carrying case, the option temperature controller for heated line (Art. No.
01G6190) is selected, on the rear of the housing of the CSS-VC.. there is a 7-pin connector for
connection of heated lines type PSP and PSP 4M and PSP 4M-W. These have at the point of
connection and termination a 7-pin connector and a 7-pin socket for connection to the portable gas
conditioning unit CSS-VC.. and the portable gas sampling probe type PSP4000H. This is the power
supply to the heated line and gas sampling probe using the CSS-VC...
3 Mains
probe
1 Mains L
2 Mains N
4 Mains
probe
6 PT100
5 PT100
Figure 10 Heated line connection for portable version to 10A
12.4.4 CONNECTING THE OPTIONAL EXPANSION MODULES
The connection of expansion modules to the CSS-VC.. is via the communication module with the
appropriate cable to the RJ-45 socket in the connection panel (see Figure 3).
The electrical connection of the individual expansion modules is described in Section 15.
NOTICE!
The RJ45 socket is only suitable for the connection of the
communication module. A network connection is not possible with it!
13 COMMISSIONING
Before commissioning, the plant and process-specific safety measures must to be observed.
Before switching on the power supply, it should be checked again that the operating voltage (see
nameplate) and the power supply voltage correspond!
W A R N U N G !
Before comissioning exclude that for one channel as well
internal as external a PT100 is connected.
Zobrazit stránku 23