Gas sampling and gas conditioning technology 4-1.1-ME
The following chart describes the functions of the operation and control board.
control of the CSS with operation and control board;
link between pin 1 and 9 (connector X2, s. 10.2.2);
control power supply;
External Control
(extern)
external control with Sub-D-Plug X2 (s. 10.2.3),
switches of the operation and control board out of
function; control power supply after switching the CSS
on ;
internal control activated;
no alarm: CSS ready for operation;
alarm : cooler-/temperature-controller alarm;
flow alarm;
liquid alarm.
no alarm: CSS ready for operation;
alarm : CSS not ready for operation,
temperature of cooler <2°C or >8°C,
or optional temperature-controller:
controller not in range
no alarm: CSS in operation;
alarm : no gas flow (i.e. inlet or outlet
is blocked),
sample gas pump not in operation,
liquid alarm,
cooler-/temperature-controller alarm;
no alarm: CSS ready for operation
alarm : condensate alarm;
CSS in sample mode,
signal contact available;
CSS in test mode,
signal contact available;
3-way solenoid valve open for sample mode;
3-way solenoid valve open for test mode;
2-way solenoid valve open for Zero Gas;
2-way solenoid valve open for Span Gas;
* LED display also for external control