Gas sampling and gas conditioning technology 4-1.1-ME
Function of the controller Fnct, factory setting = 0 : fixed-setpoint controller. Other values are
not adequate for the operation of M&C products.
Sensor type SenS, factory setting = 2: Resistance thermometer in 2-wire circuit
1: Resistance thermometer in 3-wire circuit
2: Resistance thermometer in 2-wire circuit
4: Thermocouple
Linearization Lin, factory setting = 1, Pt100
1: Pt100
9: Fe-CuNi J
11: Fe-CuNi L
12: NiCr-Ni K
Further information is in the separate manual 2-5.1.1ME of the controller 70304. The manual is avail-
able for download on the M&C-website www.mc-techgroup.com.
13 CHANGE OF PARAMETERS
To change parameters the level inhibit on the user level has to be removed.
W A R NIN G !
Observe the maximum temperature of the device to be controlled
to avoid damaging the same.
13.1 REMOVING AND ACTIVATING THE LEVEL INHIBIT
To remove the level inhibit, act as follows:
Standard display (below setpoint, up actual value ) has to be visible
Press key PGM and simultaneously for 5sec.,
display = Code 3 (all levels are locked)
Press PGM
Change value from 3 to 2 with key
The value is blinking after 2sec. and the change is taken over
The user level is unlocked now
Press EXIT
To activate the level inhibit, act as follows:
Standard display (below setpoint, up actual value ) has to be visible
Press key PGM and simultaneously for 5sec.,
display = Code 2 (all levels are locked)
Press PGM
Change value from 2 to 3 with key
The value is blinking after 2sec. and the change is taken over
The user level is locked now
Press EXIT