Gas sampling and gas conditioning technology 4-1.1-ME
Function of the controller Fnct, factory setting = 0 : fixed-setpoint controller. Other values are
not adequate for the operation of M&C products.
Sensor type SenS, factory setting = 2: Resistance thermometer in 2-wire circuit
1: Resistance thermometer in 3-wire circuit
2: Resistance thermometer in 2-wire circuit
4: Thermocouple
Linearization Lin, factory setting = 1, Pt100
1: Pt100
9: Fe-CuNi J
11: Fe-CuNi L
12: NiCr-Ni K
Further information is in the separate manual 2-5.1.1ME of the controller 70304. The manual is avail-
able for download on the M&C-website www.mc-techgroup.com.
13 CHANGE OF PARAMETERS
To change parameters the level inhibit on the user level has to be removed.
Observe the maximum temperature of the device to be controlled
to avoid damaging the same.
13.1 REMOVING AND ACTIVATING THE LEVEL INHIBIT
To remove the level inhibit, act as follows:
Standard display (below setpoint, up actual value ) has to be visible
Press key PGM and simultaneously for 5sec.,
display = Code 3 (all levels are locked)
Press PGM
Change value from 3 to 2 with key
The value is blinking after 2sec. and the change is taken over
The user level is unlocked now
Press EXIT
To activate the level inhibit, act as follows:
Standard display (below setpoint, up actual value ) has to be visible
Press key PGM and simultaneously for 5sec.,
display = Code 2 (all levels are locked)
Press PGM
Change value from 2 to 3 with key
The value is blinking after 2sec. and the change is taken over
The user level is locked now
Press EXIT