Gas sampling and gas conditioning technology 4-1.1-ME
7 TECHNICAL DATA
CSS(/C) : max. 140l/h
CSS-2 : max. 2 x 140l/h
CSS-3(/C) : max. 350l/h
CSS(/C), CSS-2: 2 x with needle valve, adjustable to 70l/h,
flow meter FM1 with flow alarm sensor
CSS-3(/C) : 2 x with needle valve, adjustable to 250l/h,
flow meter FM1 with flow alarm sensor
Sample inlet temperature**
range of adjustment: +2 °C ..... +15 °C, factory setting: +5 °C
at constant conditions < ± 0,1 °C
Glass fiber, retention rate 99,99% for particles > 0,1µm
19"-rack housing mounting 6 U, depth 350mm
CSS(/C) : approx. 15kg
CSS-2, CSS-3(/C): approx. 16,5kg
230V 50/60Hz or 115V 60Hz, CSS : 150VA
CSS-2/3: 250VA
Power terminals max. 4 mm
2
(4 x PG 13,5)
Alarm-/Control signal 15 pin Sub-D-Connector
Material of sample contacting
parts
PVDF, glass, Viton
, Novoprene
, PTFE
Measure/Check, cooler temperature, liquid alarm, flow alarm:
potential free change over contact, max. 24V/1A
solenoid valve actuated by manual or external switch
Option:
Electronic PID temperature con-
troller for heated sample lines
front panel mounting
control range : 0°C to 200°C
sensor inlet : PT 100 and Fe-CuNi
controlling outlet : 10A solid state relay
#
status alarm : integrated into the status signal
parameter : free adjustable
Electrical equipment standard
EN 60204-1 (DIN VDE 0113 Teil 1/02.86)
* other connections on request
** Maximum values in technical datas must be rated in consideration of total cooling capacity at 25 °C ambient temperature and an outlet dew
point of 5 °C.
#
standard for max. 20m heated sample line at 110W/m